back to product overview

Analytical sample gas lines with RSL fitting

H 300 series350 °C

Application potential: Maintaining the temperature of motor exhaust, CO2 measurements, industrial exhaust gasses, blast furnace exhaust gasses, air testing, environmental testing, etc.

This heated sample-extraction line prevents condensation from forming or a temperature drop below the dew point, thus no gas components are eliminated or lost.

Operating temperature 100°C / 200°C / 250°C / 350°C
Rated voltage 230 V AC/DC (other voltages up to 500 V)
Power rating Watt/metre, see type codes
Inner core DN 4 – 12 mm up to 250°C PTFE or PFA above 250°C stainless steel see Inner cores analytics
Connection fitting RSL 1.4571 steel, for cutting ring screw, without transition, see table
Heating heating conductor, structure according to DIN, moisture-proof with PE conductor braiding; > 250°C not moisture-proof
Thermal insulation depending on the operating temperature heat stabilized, close-pore silicone foam or thermal fleece, elastomer foam
Outer protective braiding polyamide black, options - see Outer protection
Hose end caps PA hard cap or elastomer cap
Temperature sensor Fe-CuNi type J, NiCr-Ni type K, PT 100 and integral control system (HTI) possible
Connection cable 3 m
Plug connection according to specification
Production lengths up to 100 m
Protection type IP65 (EN 60529), protection class I
Tolerance ±10°C

powered by webEdition CMS